11 Technical specifications
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
91
Setting
Program
Moderate COAG Dessication 4
0-6, 9, 12, 14-16, 17, 24-99 (socket 1+2),
23 (only socket 1)
ARC350MCD4_SE_091217_1
0
20
40
60
80
100
120
140
0
0
0
0
1
0
0
0
1
0
0
1
0
1
R [Ω]
P [W]
P = 120 W
P = 60 W
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “Moderate COAG
Dessication 4” = 120 W/60 W
ARC350MCD1_SE_080124_2
0
20
40
60
80
100
120
0
20
40
60
80
100
120
displayed value
P [W]
•
Diagram of power output P [W] as a
function of the “Moderate COAG
Dessication 4” setting. Measured load
resistance = 500
Ω
ARC350MCD4_SE_091217_3
0
200
400
600
800
1000
1200
1400
1600
1800
0
20
40
60
80
100
120
displayed value
U [Vp]
•
Diagram of HF output voltage U [Vp] as
a function of the “Moderate COAG
Dessication 4” setting (idle mode)