8 Preparation
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
69
8. Preparation
Pos: 43 /679-BOWA/ARC 300/350/9 Aufbereitung/9.1 Aufbereitung des Zubehörs/Desinfizieren und Reinigung @ 4\mod_1261040694772_6.doc @ 35185 @ 22
8.1.
Preparation of the accessories
X
Prepare the accessories (e.g. surgical handles, instruments, active
electrodes, neutral electrodes and cables) as described in the
corresponding operating manuals.
X
Check the accessories before and after use for damage and to
ensure that they are working properly.
8.2.
Disinfection and cleaning
NOTE
Incorrect handling of the HF device can cause damage to the unit!
X
Never sterilize the HF devices ARC 300/350. Clean or disinfect
them instead.
WARNING
Risk of electric shock and fire!
X
Unplug the power connection before cleaning the device.
X
For cleaning surfaces, use the approved cleaning
agents/disinfectants only as specified by the manufacturer.
X
Ensure that no liquid penetrates the device.
X
Ensure that the AUTOSTART function is deactivated.
1. Apply the cleaning agent and disinfectant.
2. Wipe the agent off with a sponge moistened with clean water or
with a cloth.
3. Dry the device using a clean, lint-free cloth.
Pos: 44 /679-BOWA/ARC 300/350/10 Wartung/10 Wartung/Reparatur @ 4\mod_1261040698397_6.doc @ 35204 @ 1
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...