6 Operation
54
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
6.3.22.
Program 21 "GastroCUT Pol" (optional for ARC 300)
This program is used in gastroenterology for the removal of polyps with
polypectomy snares via flexible endoscopes.
In this program, polypectomy snares are used for cutting and
coagulation. Arc control technology generates the cutting effect with
simultaneously minimized power output.
ARC control facilitates immediate cutting and prevents adhesion of the
electrodes.
The degree of coagulation increases with an increasing blend factor.
Three modes are available in this program: fast, medium, and slow.
For further information on the “GastroCut” mode, see Chapter
"GastroCUT" mode, Page 27.
;
The key/display for “Monopolar output 2”
39
and “Bipolar output 4”
48
are preconfigured.
1. Connect the polypectomy snares to the socket for “Monopolar
output 2”
40
(use an adapter if necessary).
2. Connect the foot switch to the HF device.
3. Activate the HF current via the yellow key of the foot switch for
cutting.
–or–
Activate the HF current via the blue key of the foot switch for
coagulation.
4. Set the coagulation level using the keys for scabbing level
5/6
.
5. In the case of hemorrhages:
Press the key/display for “Forced Coagulation” current
13
or use a
clip.
6. If necessary, remove the polypectomy snares and connect the
argon probe with the argon unit ARC PLUS.
7. Press the key/display for “Spray Coagulation” current
14
.
8. Press the blue key of the foot switch for coagulation.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...