6 Operation
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
59
6.4.4.
Menu program 3 "Moderate Coag Mode"
For further information on the “Moderate Coag” mode, see Chapter
"Moderate Coag" mode, Page 25.
In the “Moderate Coag Mode” menu, it is possible to change the mode
for moderate coagulation.
The following modes are available in menu program 3:
•
Standard
mode
•
Dessication 1, 2, and 3
1. Press the power limitation key for monopolar cutting
4
.
2. Press the key for scabbing level
5/6
to select the desired mode.
ª
The selected mode is saved and displayed for 3 seconds on the
bipolar coag display
29
after switching on the device.
3. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
6.4.5.
Menu program 4 "Bip Coag Mode"
In the "Bipolar Coag Mode" menu you can change the mode of forced
coagulation.
The following modes are available in menu program 4:
•
Standard COAG Mode
•
Soft COAG Mode
•
AUTOSTART is not available in "Soft COAG" mode.
•
For further information on the “Soft COAG Mode”, see
Chapter "Soft COAG" mode, Page 28.
1. Press the power limitation key for monopolar cutting
4
.
2. Press the key for scabbing level
5/6
to select the desired mode.
ª
The selected mode is saved and displayed for 3 seconds on the
bipolar coag display
29
after switching on the device.
3. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...