11 Technical specifications
84
ARC 300 / ARC 350 Operating manual
900-350_IFU_V2.6_10296-S13-20100111-EN
Setting
Program
GastroCUT Pap
22
ARC350GCPap_SE_10011_1
0
10
20
30
40
50
60
70
10
100
1000
10000
R [Ohm]
P [W]
Coag Phase Blend 9
Coag Phase Blend 8
Coag Phase Blend 7
Coag Phase Blend 6
Coag Phase Blend 5
Coag Phase Blend 4
Coag Phase Blend 2,3
Coag Phase Blend 0,1
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “GastroCUT Pap”
“Blend 0-9” during the Coag phase
ARC350GCPap_SE_100111_3
0
100
200
300
400
500
600
10
100
1000
10000
R [Ohm]
P [W]
Measurement of resistances
•
Diagram of power output P [W] as a
function of the load resistance R [
Ω
] for
the setting “GastroCUT Pap” during the
Cut phase
ARC350GCPap_SE_100111_2
0
10
20
30
40
50
60
70
0
1
2
3
4
5
6
7
8
9
Blend
P [W]
Coag Phase
Cut Phase = 520 W
•
Diagram of power output P [W] as a
function of the “Blend” setting for
“GastroCUT Pap”
Blend
1000
1000
1000
1000
1000
1000
1000
1000
1000
1000
•
Diagram of HF output voltage U [Vp] as
a function of the “GastroCUT Pap”
“Blend 0-9” (idle mode)
The short-term peak power exceeds 400 W. However, this is still
permissible since the mean value from 1 second is below 400 W.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...