6 Operation
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
45
6.3.4.
Program 3 “Abdominal Ligation”
(ligation only for ARC 350L)
•
Conventional instruments (such as handles) must not be
used in the ligation mode.
Ligation in Program 3 is only possible with “Bipolar
output 3”
43
.
•
If the software option “Ligation” is unavailable, this program
provides the function “Standard Bipolar”.
This program is used in a wide range of surgical applications, e.g.
gynecology and laparoscopy.
It facilitates permanent blood vessel sealing with the manual
instruments of the BOWA NightKNIFE
®
, LIGATOR
®
, and TissueSeal
®
series for large veins, arteries or tissue bundles.
Intended for:
•
large-lumen veins, arteries or tissue bundles
•
use in combination with instruments of the
BOWA NightKNIFE
®
, LIGATOR
®
, and TissueSeal
®
series
Not intended for:
•
tubal sterilization or sterilization by means of tubal
coagulation
•
dirty or defective electrode tips
The performance output is preconfigured and is automatically adjusted.
The sealing effect can be altered via the key for scabbing level
25
or
26
.
The set effect is displayed from 1 to 4 in the 7-segment display “Bipolar
Coag”
29
:
•
Effect 1: For a bipolar coagulation of thin structures with the
aid of a BOWA ligation instrument.
•
Effect 2: The sealing of thin blood vessels and structures with
the aid of a BOWA ligation instrument.
•
Effect 3: The sealing of medium-sized blood vessels and
structures with the aid of a BOWA ligation instrument.
•
Effect 4: The sealing of thick blood vessels and structures of
up to 7 mm with the aid of a BOWA ligation instrument.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...