6 Operation
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
61
6.4.9.
Menu program 7 "Hide Fix Prog"
Standard programs that are not required are hidden in the menu
program "Hide Fix Prog".
1. Press the power limitation key for monopolar cutting
4
.
2. Press the power limitation key for bipolar coagulation 31
to
display
all standard programs.
ª
“All fixpgm active” appears on the main display
15
.
3. Press the power limitation key for bipolar coagulation
30
to hide all
standard programs.
ª
“No fixpgm active” appears on the main display
15
.
–or–
Press the key/display for AUTOSTART
32
to show/hide individual
standard programs.
4. Press program key
16/17
to browse the standard programs.
5. Press the power limitation key for monopolar coagulation
10
to hide
the standard program.
–or–
Press the power limitation key for monopolar coagulation
11
to
show the standard program.
6. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
6.4.10.
Menu program 8 "Autostart Delay"
In the menu program “Autostart Delay”, the start-up delay period and
post-run time are changed in the AUTOSTART mode.
1. Press the power limitation key for monopolar cutting
4
.
2. Press the power limitation key for monopolar coagulation
10/11
to
adjust the start-up delay in 50 ms increments
(from 50-2500 ms).
3. Press the power limitation key for bipolar coagulation
30/31
to
adjust the post-run time in 100 ms increments (from 0-2500 ms).
4. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...