6.4.6.
Menu program 5 "Forced Coag Mode"
In "Forced Coag Mode", set the cutting/coagulation specifications for the
"Forced Coag" current.
For further information on the “Forced Coag Mode”, see Chapter "Forced
Coag" mode, Page 26.
1. Press the power limitation key for monopolar cutting
4
.
2. Press the key for scabbing level
5/6
to change the forced
coagulation mode.
3. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
6.4.7.
Menu program 5a “GastroCut Mode”
For further information on the “GastroCut Mode”, see Chapter
"GastroCUT" mode, Page 27.
1. Press the power limitation key for monopolar cutting
4
.
2. Press the key for scabbing level
5
/
6
to toggle between slow
"GastroCUT" (slow, non-accelerated), dynamic "GastroCUT"
(default, medium) and fast dynamic "GastroCUT" (accelerated,
fast).
3. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
6.4.8.
Menu program 6 "Show Prev Inf-No"
The errors saved in the HF device are called up in the menu program
"Show Prev Inf-No".
1. Press the power limitation key for monopolar cutting
4
.
2. Press the key for scabbing level
5
to call up the individual error.
ª
The error number will appear on the main display
15
.
3. Press the key/display for AUTOSTART
32
to call up the last 10
errors.
ª
The error number will appear on the main display
15
.
4. Press the power limitation key for monopolar cutting
3
to return to
the menu programs.
The service manual contains further information on menu program 6.
Содержание ARC 300
Страница 1: ...OPERATING MANUAL HIGH FREQUENCY SURGICAL DEVICE...
Страница 2: ......
Страница 4: ...4 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...
Страница 5: ...900 350_IFU_V2 6_10296 S13 20100111 EN ARC 300 ARC 350 Operating manual 5...
Страница 8: ...Legend 8 ARC 300 ARC 350 Operating manual 900 350_IFU_V2 6_10296 S13 20100111 EN...