7 Detecting and correcting errors
900-350_IFU_V2.6_10296-S13-20100111-EN
ARC 300 / ARC 350 Operating manual
67
Error no.
(INF)
Signa-
ls/System
events
Cause
Corrective measures
149
•
3-pin plug of the
handpiece has been
inserted incorrectly
•
Handpiece possibly
defective
•
Monopolar 4 mm
connection cable has
been connected to output
2
40
in the right-hand 4
mm connection socket
X
Check the handpiece.
X
Replace the handpiece.
X
Connect the 4 mm connection cable
to output 2
40
in the left-hand 4 mm
connection socket.
X
Contact the service department if an error cannot be corrected with
the measures listed; see Chapter Technical service, Page 72.
Errors not listed
X
Contact the service department regarding errors that are not shown
in the error list; see Chapter Technical service, Page 72.
X
If the expected change in the tissue does not occur and no error
message appears, check the parameters and the accessory
connections.