Remote Interface Operation
The following commands show how to use the remote interface
operation to disable or enable the beeper:
SYSTem:BEEPer
SYSTem:BEEPer:STATe {OFF|ON}
4.4.3 Reading Memory (Store & Recall)
ATM3500A has a memory capacity of 2000 readings. The readings are
stored in first-in-first-out order and the memory type is volatile, which
means the stored readings will be cleared when the multimeter is
power-off. The reading memory feature can be used for all
measurement functions, math operations, trigger operations and
reading hold.
How to use the reading memory
You can store the readings and access to the stored readings through
either the front panel operation or the remote interface operation.
※
Note: Each datum stored from ATM3500A to
remote interface will be in a first in and first out
condition.
Front Panel Operation
Before using the reading memory feature, you need to
select a
measurement function
(or the
math
function) first and then select
the
trigger mode
.
To store readings, use the following steps:
Press STORE button, and the multimeter will start to store the readings
produced until the specified number of readings is reached. The default
number of readings to store is 100. To change this number, press
101