Table 3-1
(Temperature Ranges for Each Thermocouple Type)
Sensor Type
Temperature
Range(°C)
B
600~1820
C
0~2316
E
-250~1000
J
-210~1200
K
-200~1372
N
-200~1300
R
0~1767
S
0~1767
T
-250~400
3.7.1 Thermocouple Measurements
Connect the thermocouple adapter to the banana jacks on the front
panel. The difference between each type is subject to the patch
thermal leads.
※
Note: Only connection via the front panel can be used for
temperature measurements.
How to measure thermocouple
1
Use terminals switch to select terminals on the front panel.
2
To measure thermocouple, you must use a thermocouple adapter
and a sensor, (K type for example), with the low thermal patch lead
43