SAMPle:COUNt <value>
B. Number of triggers
Although the meter normally takes one trigger before returning to the
“idle” state, you can manually specify the number of triggers it accepts
before the “idle” state. However, this can only be done through the
remote interface. The following command shows how to set multiple
triggers per returning idle state.
TRIGger:COUNt <value>
C. Reading hold
The reading hold feature is used to hold a stable reading on the
display. When a reading is stable, enabling the reading hold will hold
the stable reading and trigger a beeping sound. Reading hold feature
depends on an adjustable sensitivity band to determine if a reading is
stable. The multimeter considers a reading stable when three
consecutive readings are within the sensitivity band. Chapter 4.4.6 will
explain how to adjust the sensitivity band.
How to use reading hold
This feature is only available from the front panel operations. To enable
reading hold, press both SHIFT and AUTO TRIGGER buttons. To disable
it, simply press AUTO TRIGGER. The locations of the buttons are
shown with red rectangle frames in
Figure 4-12
.
Figure 4-12
D. Trigger delay
This feature is useful for the user who needs a longer delay time to
77
+