You can set up the thermocouple configuration either through the front
panel operation or through the remote interface operation.
Front Panel Operation
1.
Configure the thermocouple type and unit. Pressing CONFIG
>
SHIFT
>
TEMP (TCOUPL), and then using PREV and NEXT to select
the type and unit.
2.
Measure a known and accurate reference temperature (T1).
The result showing on the ATM3500A is (T2).
3.
Adjust the simulated value (23.0°C to 23.0°C + T3.)
4.
Configure the Simulated value by pressing CONFIG
>
SHIFT
>
TEMP(TCOUPL)
>
NEXT
>
SIMULATED
>
ENTER, and by
buttons
「
PREV
」
or
「
NEXT
」
and &
△ ▽
assistance to adjust the value.
Remote Interface Operation
Use the following commands to set up the thermocouple configuration.
SENSe
:UNIT {Cel|Far|K}
SENSe
:UNIT?
SENSe:
TCOuple:TYPE {E|J|K|N|R|S|T}
SENSe:
TCOuple:SIMulated {<
value
>|MINimum|MAXimum}
4.1.9 Remote Interface Selection
The multimeter supports both GPIB/RS-232 and USB interfaces, but
only one interface can be activated at a time. If you are using GPIB,
you must set the address for the multimeter. You can set the address
to any value from 0 and 31. The address is set to “22” when the
multimeter is shipped from the factory.
Note: The remote interface can be set through the front panel
※
69