average to produce a reading for dislay.filter takes the specified
number of reading conversion, averages them and produces a reading.
It then flushes the stack and starts over with an new User can set the
number to anything from 1 to 100Consequentter yields one reading for
display every specified number measurement r
Digital filter is not available for diode, continuity, frequency and period
measurements.
Default
The digital filter is enabled and is in moving average mode with 10
readings by default.
How to enable/disable digital filter
Press FILTER button to switch the digital filter function.
The “FILT” anunnciator indicates the state of the digital filter. When it
is lit, the filter is enabled.
How to configure digital filter fo
You can configure the digital Filter either through the front panel
operation or through the remote interface operation.
Front Panel Operation
For READINGS setting, please press CONFIG and then FILTER button
for selecting READINGS. Press ENTER to access. Use
◁
,
▷
,
△
and
▽
buttons to set the READINGS (from 2 to 100) in order to get an
average value. Press ENTER when it’s ready.
For MODE setting, please press CONFIG and then FILTER button for
selecting MODE. Press ENTER to access. Use
◁
,
▷
to select MOVING
AVG. or REPEAT AVG.. Please ENTER to confirm.
Press FILTER button to enable or disable this function.
Remote Interface Operation
Use the following commands to set your digital filter:
SENSe:AVERage:TCONtrol {MOVing|REPeat}
SENSe:AVERage:TCONtrol?
SENSe:AVERage:COUNt {<
value
>|MINimum|MAXimum}
SENSe:AVERage:COUNt? [MINimum|MAXimum]
56