Function
Range
Resolution
Shunt
Resistance
1 Year
(23°C ± 5°C)
DCI
(DC
Current)
10.00000mA
10 nA
5.1Ω
0.050 + 0.020
100.0000mA
100 nA
5.1Ω
0.050 + 0.005
1.000000A
1 uA
0.1Ω
0.100 + 0.010
3.00000A
10 uA
0.1Ω
0.120 + 0.020
Function
Range
Resolution
Test Current
1 Year
(23°C ± 5°C)
Resistance
4
100.0000 Ω
100 uΩ
1 mA
0.010 + 0.004
1.000000 KΩ
1 mΩ
1 mA
0.010 + 0.001
10.00000 KΩ
10 mΩ
100 uA
0.010 + 0.001
100.0000 KΩ
100 mΩ
10 uA
0.010 + 0.001
1.000000 MΩ
1 Ω
5 uA
0.010 + 0.001
10.00000 MΩ
10 Ω
500 nA
0.040 + 0.001
100.0000 MΩ
100 Ω
500 nA||10
MΩ
0.800 + 0.010
Diode Test
1.0000V
10 uV
1mA
0.010 + 0.020
Continuity
1000.00KΩ
10 mΩ
1mA
0.010 + 0.030
Frequency and Period Characteristics
Accuracy ± (% of reading)
5
Function
Range
6
Frequency (Hz)
1 Year
23º C±5º C
4
a. To use 2- / 4-Wire
Ω Measurement
, the Null math function must
be enabled.
b. Max. Lead Resistance: 10% of range per lead for 100Ω and 1KΩ
ranges; 1KΩ per lead for all other ranges.
c. Input protection: 1000 V at all ranges.
5
The Specifications are under the conditions, 6 ½ digits and 2-hour
warm-up.