TRIGger:SOURce IMMediate
C. Single Trigger Mode (accessible only through the front
panel)
Definition
Single trigger mode takes one reading (or specified number of
readings) each time when you press SINGLE key. (Please refer to 4.2.3
for setting the number of samples on each trigger.) When the TRIG
annunciator on the display is lit, the meter is ready for next trigger
event. The single trigger mode is only available through the front panel
operations.
How to use Single Trigger
First select the measurement function, and then press SINGLE key on
the front panel to set the trigger mode. The location of the button is
shown with a red rectangle frame in
Figure 4-10
.
Figure 4-10
4.2.2 Trigger Source
73
Содержание ATM3500A
Страница 1: ...ATTEN EU ATM3500A 6 5 Digit Digital Multimeter...
Страница 2: ...User s Manual ATM3500A DMM User s Manual 2 Printed date 11 2011 Version 1 05...
Страница 50: ...Figure 3 18 50 5...
Страница 120: ...668 Manual Calibration Version Error 669 Self Test Version Error 670 Command Version Error 120...