rectangle frames in
Figure 4-17
.
※
Note: Press NULL again to disable this feature. The “MATH”
anunnciator on the display indicates the state of a mathematical
feature.
Figure 4-17
The Remote Interface Operation
You can use the following commands on your PC terminal to make a
null measurement.
CALCulate:FUNCtion NULL
CALCulate:STATe {OFF|ON}
CALCulate:STATe?
CALCulate:NULL:OFFSet
{
<value>|MAXimum|MINimum
}
4.3.5 Limits Test
The limits testing operation allows you to adjust a maximal and a
minimal limit values. The multimeter beeps and an “HI” or “LO”
message will be shown when the reading exceeds the upper or lower
limit respectively. You can specify the limit values and the values are
stored in a volatile memory. The default values for both upper and
lower limits are “0”. This function is available to all except continuity
and diode measurements.
※
Note: You can also use this HI/LO measurement as a Pass/Fail
86