Frequency
&
Period
100mV
to
750V
3-5
0.10
5-10
0.05
10-40
0.03
40-300K
0.01
AC Characteristics
Accuracy ± (% of r % of range)
7
Function
Range
8
Resolution
Frequency
(Hz)
1 Year
(23°C ± 5°C)
ACV (AC
TRMS
Voltage)
9
100.0000mV
0.1 uV
3-5
1.00 + 0.04
5-10
0.35 + 0.04
10-20K
0.06 + 0.04
20-50K
0.12 + 0.05
50K – 100K
0.60 + 0.08
100K – 300K
4.00 + 0.50
1.000000V
To
750.000V
1.0 uV
To
1mV
3-5
1.00 + 0.03
5-10
0.35 + 0.03
10-20K
0.06 + 0.03
20-50K
0.12 + 0.05
50K – 100K
10
0.60 + 0.08
6
Measurement over 20% at all ranges is allowable except the range
750 VAC.
7
The Specifications are under the test conditions, 6 ½ digits and 2-
hour warm-up, slow AC filter (3 Hz Bandwidth), sine wave input.
8
Measurement over 20% at all ranges is allowable except the range
750 VAC.
9
The Specifications are under the test condition, sine wave
input>5% of range. When the inputs are within 1% to 5% of range
and<50 KHz, the error, 0.1% of range must be added. For inputs
within 50 KHz to 100 KHz, the 0.13% of range must be added.
1
0
The ACV range 750V is limited by 100 KHz.
146
Содержание ATM3500A
Страница 1: ...ATTEN EU ATM3500A 6 5 Digit Digital Multimeter...
Страница 2: ...User s Manual ATM3500A DMM User s Manual 2 Printed date 11 2011 Version 1 05...
Страница 50: ...Figure 3 18 50 5...
Страница 120: ...668 Manual Calibration Version Error 669 Self Test Version Error 670 Command Version Error 120...