CALIBRATION
11. IR test, XXX.XM
W
range
Press the
Ú
key, the program will advance to the XXX.XM
W
range calibration of
the IR test. The display will show:
XXX.XM
W
,STD 50M
W
<
T E S T
>
t o C a l i b r a t e
Connect a 50M
W
/0.25W/1000VDC standard resistor to the output connectors, then
press the TEST button, the program will automatically calibrate the XXX.X M
W
range of IR test and the display will show:
XXX.XM
W
,STD 50M
W
OK
<
T E S T
>
t o C a l i b r a t e
12. IR test, XXXXM
W
range
Press the
Ú
key, the program will advance to the XXXXM
W
range calibration of the
IR test. The display will show:
XXXXM
W
,STD 50M
W
<
T E S T
>
t o C a l i b r a t e
Connect a 50M
W
/0.25W/1000VDC standard resistor to the output connectors, then
press the TEST button, the program will automatically calibrate the XXXX M
W
range of IR test and the display will show:
XXXXM
W
,STD 50M
W
OK
<
T E S T
>
t o C a l i b r a t e
13. Ground Bond, AC Voltage (Model 7504SA only)
Press the
Ú
key, the program will advance to the AC Voltage calibration of the
Ground bond test. The display will show:
Ground V, 7V
<
T E S T
>
t o C a l i b r a t e
Connect a AC standard voltage meter which can measure up to 10VAC to the High
current and Return output leads, then press the TEST button, the program will
automatically generate an output of about 7VAC and the display will show:
G-Voltage = V
Enter Standard V-out
Use the Numeric keys to enter the standard value of voltage, unit V, and then press