CONTROLS
25
15. UP-DOWN ARROW KEYS:
(PARAMETER KEY) Use these keys to enter and move
through the function parameter menus for AC-W, DC-W, IR, and G-BOND tests.
16. DATA ENTRY KEYS:
Use these keys to input numeric parameters followed by
the ENTER key.
17. ENTER/REVIEW KEY:
Use this key as an ENTER key to accept numeric data for
parameter settings. Or use this key as a REVIEW key to recall up to 8 connected test
results after a test has been completed.
18. HIGH VOLTAGE OUTPUT JACK:
For the connection of the detachable 5 foot (1.52
m) high voltage test lead. The silicone rubber insulation is flexible for easy handling and is
rated at 30KVDC. The jack is recessed for safety when this lead is not being used.
19. HIGH VOLTAGE ON INDICATOR:
This indicator flashes to warn the operator that
high voltage is present at the high voltage output terminal.
20. RETURN OUTPUT JACK:
For the connection of the detachable 5 foot (1.52 m)
return test lead. This lead is always used when performing a test. On model 7504SA this
jack will carry high current for the Ground Bond test
21. CURRENT OUTPUT JACK: (
Model 7504SA only) For the connection of the
detachable 5 foot (1.52 m) high current output lead used for the ground bond test. This
lead is only used for the ground bond test.
GROUND CONTINUITY JACK: (
Model 7500DT only) For the connection of the
detachable 5 foot (1.52 m) low current return lead used for the ground continuity test.
1
10 11
13
12
14 15
16
17
18 19 20
21
2 3
4
5
6
7
8 9