![Associated Research HypotULTRA II 7500DT Скачать руководство пользователя страница 75](http://html1.mh-extra.com/html/associated-research/hypotultra-ii-7500dt/hypotultra-ii-7500dt_operation-and-service-manual_2989479075.webp)
OPERATION
71
No more than 15 device loads connected to each bus, with no less than two-thirds
powered on.
Note:
A bus extender which is available from numerous manufacturers is available
to overcome these limitations.
8.6 Interface Functions
The capability of a device connected to the bus is specified by its interface functions.
These functions provide the means for a device to receive, process, and send
messages over the bus. The interface functions are listed in the chart below. All
functions may be controlled over the bus except input voltage which is Selectable on
the rear panel.
GPIB INTERFACE FUNCTIONS
IEEE-488
INTERFACE
Complete
handshake
capability
Talker/Listener functions
Service request capability
No remote/local capability
No parallel poll capability
No device clear capability
No device trigger capability
No controller capability
3 state driver
CONTROLLABLE
ITEMS
Test/Reset
control
Setting of test status/parameters for test
Display Reading
Results Reading
DATA CODES
ASCII
DELIMITER
CR + LF (+ EOI)
8.7 GPIB Address
Each device on the GPIB (IEEE-488) interface must have a unique address. You
can set the address of the 7504SA or 7500DT to any value between 0 and 30. The
address is set to 8 when the instrument is shipped from the factory. The address can
only be set from the front panel. The address is stored in non-volatile memory and
does not change when the power has been off or after a remote interface reset.
8.8 RS-232 Interface
This interface may be substituted for the standard GPIB interface on model 7500DT
or 7504SA. This interface provides all of the function control of the GPIB interface
with the exception of the SRQ functions. All commands can be found in the
command list portion of the Operation section 8.9 of this manual.