OPERATION
44
2.2.10 Step Connection setting
Advance the menu to the Connection parameter. The display will show:
Connect = ON
<
ENTER
>
to Select
or
Connect =
OFF
<
ENTER
>
to Select
Use the ENTER key to select the Step Connection mode, then press the
Ù
or
Ú
key
to advance the program to another test parameter or press the EXIT key to exit
from the setting mode to operation mode.
If Connect is set to ON, the next step in the sequence will be executed. If Connect is
set to OFF, the test sequence will stop at this step.
If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
2.3 Insulation Resistance test setting
Press the INSULATION key and the display will show:
IR Set XXX.X s
MXX-X XXXX V XXXXX M
W
Note: X = the numeric
IR Set
: Insulation Resistance test setting screen
XXX.X s : Delay Time setting
MXX
:
Memory
Program
number
-X
: Test Step number
XXXX V : DC Output Voltage
XXXX M
W
: Low-Limit of Insulation Resistance
Use the
Ù
or
Ú
arrow keys to progress through the test parameters menu. The
Ú
key will advance forward and
Ù
key will advance backward. The sequential
forward menu items are Voltage, Charge-LO, HI-Limit, LO-Limit, Delay Time,
Scanner Set, Connect.
2.3.1 Output Voltage setting
Advance the menu to the Voltage parameter. The display will show:
Voltage = XXXX V
Range : 0 - 6000
Use the Numeric keys to enter the voltage setting, then press the ENTER key. The
program will store the voltage setting and advance to the Charge-LO parameter
automatically. The unit is volt and 1 volt per step. Press the EXIT key to exit from
the setting mode to the operation mode if all parameters have been set.