OPTIONS
11 Remote Interlock
Option 07 is a Remote Interlock feature that inverts the present reset logic on the
instrument from normally opened (must close contacts to activate the reset) to
normally closed (must open contacts to activate the reset). Hardware and software
have been reconfigured to provide the interlock connections on pins 4 and 5 of the
Remote Interface, Signal Input port. This reset scheme is designed for use with an
external safety interlock device that utilizes a "Fail-When-Open" configuration on its
output interface. The instrument can still be used without the external reset device as
long as the Interlock Connector 38037 (provided with unit) is plugged into the
Remote Interface, Signal Input port. If there is nothing connected to the Remote
Interface, Signal Input port to provide a connection to the interlock (reset), then the
instrument will not function. Please refer to page 65 for further information about
the remote interface.
12 Enhanced Ramping
AC Ramp Down
The AC ramp down feature allows the AC voltage applied to the DUT, to be
reduced linearly over a specified period at the end of a hipot test. There are three set
up screens associated with this feature that will allow control of Ramp-Up, Dwell,
and Ramp-Down.
The setup screens will appear as follows:
Ramp Up = XXX.X s
Range : 0.1 - 999.9
Use the Numeric keys to enter the Ramp Up time setting, then press the ENTER
key. The program will store the Ramp Up Time setting and advance to the Dwell
Time parameter automatically. The unit is second and 0.1 second per step.
Dwell Time= XXX.X s
Range : 0.3 - 999.9 0=Constant
Use the Numeric keys to enter the Dwell time setting, then press the ENTER
key. The program will store the Dwell Time setting and advance to the Ramp Down
Time parameter automatically. The unit is second and 0.1 second per step.
Ramp Down = XXX.X s
Range : 0.0 - 999.9
Use the Numeric keys to enter the Ramp Down time setting, then press the
ENTER key. The program will store the Ramp Down Time setting and advance
to the Frequency parameter automatically. The unit is second and 0.1 second per
step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.