OPERATION
43
Fail mode selection automatically. The display will show:
Arc Fail = ON
<
ENTER
>
to Select
or
Arc Fail =
OFF
<
ENTER
>
to Select
Use the ENTER key to select the Arc Fail mode, then press the
Ù
or
Ú
key to
advance the program to another test parameter or press the EXIT key to exit from
the setting mode to the operation mode.
If the Arc Fail mode is set to ON, the program will indicate an arc failure when the
arc current is over the setting. The program will shut down the test immediately and
the ARC indicator on the front panel will light.
If the Arc Fail mode is set to OFF, the program will not indicate an arc failure when
the arc current is over the setting. The program will not shut down the test but the
ARC indicator on the front panel will light only when arcing is present.
2.2.9 Scanner Channel setting
Advance the menu to the Scanner Set parameter. The display will show:
Scanner Set CH = 1 - 16
XXXXXXXXXXXXXXXX
Note: X = H, O or L.
Use the Numeric keys to enter the status of High, Open or Low for each scanner
channel, then press the ENTER key. The program will store the scanner channel
status and advance to the Connection setting automatically. The following table is
the cross reference of the numeric keys, which are used for the scanner channel
setting.
Channel
Status
Numeric
Key
Description
H (High)
1
Connected to High Voltage Output
L (Low)
0
Connected to Return
O (Open)
·
Not connected to High Voltage or
Return
The Channel setting will start from Channel 1 on the left end and stop at Channel 16
on the right end. The program will set any unassigned channel(s) to OPEN
automatically after pressing the ENTER key. Therefore it is necessary to set only
the channels that are required for the test.
If the external Scanner is not connected, this setting is not necessary.