OPERATION
36
2. Function Parameters Setup Procedures
Before going to setup the Test Parameters, make sure that the Keyboard is in the
Unlock mode, then follow this procedures to setup the Test Parameters.
2.1 AC Withstand Voltage test setting
Press the ACW key and the display will show:
ACW Set XXX.X s
MXX-X X.XX KV XX.XX mA
Note: X = Numeric
ACW Set : AC Withstand Voltage test setting screen
XXX.X s : Dwell Time setting
MXX
:
Memory
Program
number
-
X
:
Test
Step
number
X.XX KV : AC Output Voltage setting
XX.XX mA : AC High-Limit current setting
Use the
Ù
or
Ú
arrow keys to progress through the test parameters menu. The
Ú
key will advance forward and
Ù
key will advance backward. The sequential
forward menu items are Voltage, HI-Limit, LO-Limit, Ramp Time, Dwell Time,
Frequency, Arc Sense, Arc Fail, Scanner Set, Offset, Connect.
2.1.1 AC Output Voltage setting
Advance the menu to the Voltage parameter. The display will show:
Voltage = XXXX V
Range : 0 - 5000
Use the Numeric keys to enter the voltage setting, then press the ENTER key. The
program will store the voltage setting and advance to the High Limit parameter
automatically. The unit is volt and 1 volt per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
2.1.2 HI-Limit Current setting
Advance the menu to the HI-Limit parameter. The display will show:
HI-Limit = XX.XX mA
Range : 0.00 - 99.99
Use the Numeric keys to enter the HI-Limit setting, then press the ENTER key. The
program will store the HI-Limit setting and advance to the LO-Limit parameter
automatically. The unit is mA and 0.01 mA per step. Press the EXIT key to exit
from the setting mode to the operation mode if all parameters have been set.