OPERATION
39
status and advance to the Connection setting automatically. The following table is
the cross reference of the numeric keys, which are used for the Scanner channel
setting.
Channel
Status
Numeric
Key
Description
H (High)
1
Connected to High Voltage Output
L (Low)
0
Connected to Return
O (Open)
·
Not connected to High Voltage or
Return
The Channel setting will start from Channel 1 on the left end and stop at Channel 16
on the right end. The program will set any unassigned channel(s) to OPEN
automatically after pressing the ENTER key. Therefore it is necessary to set only
the channels that are required for the test.
If the external Scanner is not connected, this setting is not necessary.
2.1.9 Step Connection setting
Advance the menu to the Connection parameter. The display will show:
Connect = O N
<
ENTER
>
to Select
or
Connect =
O F F
<
ENTER
>
to Select
Use the ENTER key to select the Step Connection mode, then press
Ù
or
Ú
key to
advance the program to another test parameter or press the EXIT key to exit from
the setting mode to the operation mode.
If Connect is set to ON, the next step in the sequence will be executed. If Connect is
set to OFF, the test sequence will stop at this step.
If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
2.2 DC Withstand Voltage test setting
Press the DCW key and the display will show:
DCW Set XXX.X s
MXX-X X.XX KV XXXXX
m
A Note: X = Numeric
DCW Set : DC Withstand Voltage test Setting screen
XXX.X s : Dwell Time setting
MXX
:
Memory
Program
number
-
X
:
Test
Step
number
X.XX KV : DC Output Voltage setting
XXXX
m
A : DC High-Limit current setting