OPTIONS
85
DCW
Dwell
XXX.X s
MXX-X - . - - KV - - - -
m
A
At the end of the dwell cycle when the voltage begins to ramp down, the display will
show the values being updated in real time as follows:
DCW Ramp-down XXX.X s
MXX-X X.XX KV XXXX
m
A
NOTE:
In order to see the DCW values measured during the dwell cycle the review
key must be pressed.
IR Ramp Down
The IR ramp down feature allows the IR voltage applied to the DUT, to be reduced
linearly over a specified period at the end of an insulation resistance test.
There are three set up screens associated with this feature that will allow control of
Ramp-Up, Dwell, and Ramp-Down.
The setup screens will appear as follows:
Ramp Up = XXX.X s
Range : 0.1 - 999.9
Use the Numeric keys to enter the Ramp Up time setting, then press the ENTER
key. The program will store the Ramp Up Time setting and advance to the Dwell
Time parameter automatically. The unit is second and 0.1 second per step.
Delay Time= XXX.X s
Range : 0.5 - 999.9 0=Constant
Use the Numeric keys to enter the Dwell time setting, then press the ENTER
key. The program will store the Dwell Time setting and advance to the Ramp Down
Time parameter automatically. The unit is second and 0.1 second per step.
Ramp Down = XXX.X s
Range : 1.0 - 999.9 0=OFF
Use the Numeric keys to enter the Ramp Down time setting, then press the
ENTER key. The program will store the Ramp Down Time setting and advance
to the Frequency parameter automatically. The unit is second and 0.1 second per
step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.