CALIBRATION
91
5. IR test, DC Voltage
Press the
Ú
key, the program will advance to the DC Voltage calibration of the IR
test. The display will show:
IR Voltage, 1000V
<
T E S T
>
t o C a l i b r a t e
Connect a DC standard voltage meter which can measure up to 1000V to the output
connectors and then press the TEST button, the program will automatically generate
an output of about 1000VDC and the display will show:
IR-Voltage = V
Enter Standard V-out
Use the Numeric keys to enter the standard value of DC voltage, unit V, and then
press the ENTER key to store the standard value of IR test voltage for calibration
and the display will show:
IR Voltage, 1000V OK
<
T E S T
>
t o C a l i b r a t e
6. AC Hipot, 100mA current range
Press the
Ú
key, the program will advance to the AC 100mA range calibration of the
hipot test. The display will show:
AC 100mA, 10K
W
<
T E S T
>
t o C a l i b r a t e
Connect a resistor about 10K
W
/100W in series with a AC standard Ammeter which
can measure up to 100mA to the output leads. The Ammeter should be connected
to the return lead, then press the TEST button, the program will automatically
generate an output of about 1000V/100mA and the display will show:
Current = mA
Enter Standard I-out
Use the Numeric keys to enter the standard value of current, unit mA, and then
press the ENTER key to store the standard value of AC 100mA range for
calibration and the display will show:
AC 100mA, 10K
W
OK
<
T E S T
>
t o C a l i b r a t e