INSTALLATION
28
INSTALLATION
Introduction
This section contains information for the unpacking, inspection, preparation for use and
storage of your Associated Research, Inc., product.
Unpacking and Inspection
Your instrument was shipped in a custom foam insulated container that complies with
ASTM D4169-92a Assurance Level II Distribution Cycle 13 Performance Test Sequence.
If the shipping carton is damaged, inspect the contents for visible damage such as dents,
scratches or broken meters. If the instrument is damaged, notify the carrier and the
Associated Research customer support department immediately. Please save the shipping
carton and packing material for the carriers inspection. Our customer support department
will assist you in the repair or replacement of your instrument. Please do not return your
product without first notifying us and receiving an RMA (return material authorization)
number.
Preparation for Use
Power Requirements and Line Voltage Selection
This instrument requires a power source of either 100, 115, 200 or 230 volts AC ± 10%,
47-63 Hz. Please check the rear panel to be sure the proper switch setting is selected for
your line voltage requirements before turning your instrument on.
Do not switch the line voltage selector switches located on the rear panel
while the instrument is on or operating. This may cause internal damage
and represents a safety risk to the operator.
Power Cable
BEFORE CONNECTING POWER TO THIS INSTRUMENT, THE
PROTECTIVE GROUND (EARTH) TERMINALS OF THIS INSTRUMENT
MUST BE CONNECTED TO THE PROTECTIVE CONDUCTOR OF THE LINE (MAINS)
POWER CORD. THE MAIN PLUG SHALL ONLY BE INSERTED IN A SOCKET OUTLET
(RECEPTACLE) PROVIDED WITH A PROTECTIVE GROUND (EARTH) CONTACT. THIS
PROTECTIVE GROUND (EARTH) MUST NOT BE DEFEATED BY THE USE OF AN
EXTENSION CORD (POWER CABLE) WITHOUT A PROTECTIVE CONDUCTOR
(GROUNDING).
CAUTION
WARNING