![Associated Research HypotULTRA II 7500DT Скачать руководство пользователя страница 77](http://html1.mh-extra.com/html/associated-research/hypotultra-ii-7500dt/hypotultra-ii-7500dt_operation-and-service-manual_2989479077.webp)
OPERATION
73
8.9 Interface Bus Command List
The following commands are used to toggle ON/OFF functions or menu selection
items. No other values or parameters are needed to execute these commands.
However when using commands that are usable for a certain function only (function
specific), the appropriate function select command FC, FD, FE, or FF must be
executed first, to access the parameter for that function.
Command Function
Function
Specific
FA Test
FB
Reset
FC
ACW Function Select
FD
DCW
Function
Select
FE
IR Function Select
FF
GND Function Select (7504SA only)
FG
ACW/DCW:
Continuity
On
(7500DT
Only)
x
FH
ACW/DCW:
Continuity
Off
(7500DT
Only)
x
FI
ACW: 60Hz
x
FJ
ACW: 50Hz
x
FK
ACW:
Arc
Fail
On
x
FL
ACW: Arc Fail Off
x
FM
DCW:
Arc
Fail
On
x
FN
DCW:
Arc
Fail
Off
x
FO
GND:
60hz
(7504SA
only)
x
FP
GND: 50hz (7504SA only)
x
FQ
Connect
On
FR
Connect Off
FS
Fail Stop On
FT
Fail Stop Off
FU
GND: Auto Offset
(7504SA only)
x
FV
DCW:
Auto
Charge-LO
x
FW
IR:
Auto
Charge-LO
x
F0
Enable All Pass SRQ
F1
Disable All Pass SRQ
F2
Enable Fail SRQ
F3
Disable Fail SRQ
F4
Enable Abort SRQ
F5
Disable Abort SRQ
F6
Enable Error Command SRQ
F7
Disable Error Command SRQ
F8
DCW: Ramp-HI On
x
F9
DCW: Ramp-HI Off
x