OPERATION
51
The display will show the measured m
W
Offset value and the program will use this
value for the Offset setting. Each step has its own individual Offset value and must
be set separately. This allows the user to compensate for different lead lengths when
using a scanner or external fixturing for Ground Bond testing.
Then press the
Ù
or
Ú
key to advance the program to another test parameter setting
or press the EXIT key to exit from the setting mode to operation mode.
2.4.9 Step Connection setting
Advance the menu to the Connection parameter. The display will show:
Connect = ON
<
ENTER
>
to Select
or
Connect =
OFF
<
ENTER
>
to Select
Use the ENTER key to select the Step Connection mode, then press the
Ù
or
Ú
key
to advance the program to another test parameter or press the EXIT key to exit
from the setting mode to the operation mode.
If Connect is set to ON, the next step in the sequence will be executed. If Connect is
set to OFF, the test sequence will stop at this step.
If Step 8 is set to ON the test process will be connected to the first step of the next
Memory.
3. System Parameter Settings
Use the SETUP key to progress through the menu of System Parameters.
Successive key presses will advance the menu forward. The sequential forward
menu items are: PLC Remote, Address, Contrast, Volume, Fail Stop.
The setting of system parameters affect the operating conditions of the instrument
and are separate from the functional settings. The system settings are also global
and are not specific to any memory location.
3.1 PLC Remote selection
Press the Setup key to advance to the PLC Remote parameter. The display will
show:
PLC Remote = ON
<
ENTER
>
to Select
or
PLC Remote =
OFF
<
ENTER
>
to Select
Use the ENTER key to select the mode PLC Remote Control.
After selecting PLC Remote mode, press the SETUP key to advance to the address
setting if the GPIB interface card is installed on this instrument. If this instrument