OPERATION
62
If the DUT current exceeds the HI-Limit of DC Withstand Voltage test and the
leakage current is within the metering range, the display will show:
DCW
HI-Limit XXX.X s
MXX-X X.XX KV XXXX
m
A
If the DUT current exceeds the HI-Limit of DC Withstand Voltage test and the
leakage current is not within the metering range, the display will show:
DCW
HI-Limit
XXX.X s
MXX-X X.XX KV
>
9999
m
A
If the DUT current is well beyond the metering range of DC Withstand Voltage test
the instrument assumes that the failure is due to a short circuit, the display will show:
DCW
Short
XXX.X s
MXX-X - . - - KV
>
9999
m
A
If the DUT current is well beyond the metering range of DC Withstand Voltage test
and an Arcing condition beyond the Arc Sense limit is indicated, the display will
show:
DCW Breakdown XXX.X s
MXX-X X.XX KV
>
9999
m
A
If the DUT current falls below the LO-Limit of DC Withstand Voltage test and the
leakage current is within the metering range, the display will show:
DCW
LO-Limit XXX.X s
MXX-X X.XX KV XXX.X
m
A
If the DUT current within the metering range of the DC Withstand Voltage test and
an Arcing current exceeds the Arc-Sense limit and the Arc functions is set to ON
then an Arc failure has occurred and the display will show:
DCW
Arc-Fail XXX.X s
MXX-X X.XX KV XX.X
m
A
If the GFI threshold is exceeded during of the DC Withstand Voltage test, the
display will show:
DCW GND-Fault XXX.X s
MXX-X X.XX KV XXXX
m
A