OPERATION
59
meter readings are taken, the display will show:
ACW Abort XXX.X s
MXX-X - . - - KV - - . - - mA
or
ACW Abort
XXX.X s
MXX-X X.XX KV - - . - - mA
At the beginning of AC Withstand Voltage test when the voltage begins to ramp but
before the meter readings are taken, the display will show:
ACW Ramp XXX.X s
MXX-X - . - - KV - - . - - mA
During the AC Withstand Voltage test when the values are being updated in real
time during the ramp cycle, the display will show:
ACW
Ramp
XXX.X s
MXX-X X.XX KV XX.XX mA
During the AC Withstand Voltage test when the values are being updated in real
time during the dwell cycle, the display will show:
ACW
Dwell
XXX.X s
MXX-X X.XX KV XX.XX mA
If the ramp time is very short and the program has not read the meter readings, the
display will show:
ACW
Dwell
XXX.X s
MXX-X - . - - KV - - . - - mA
If the DUT current exceeds the HI-Limit of AC Withstand Voltage test and the
leakage current is within the metering range, the display will show:
ACW
HI-Limit
XXX.X s
MXX-X X.XX KV XX.XX mA
If the DUT current exceeds the HI-Limit of AC Withstand Voltage test and the
leakage current is not within the metering range, the display will show:
ACW
HI-Limit XXX.X s
MXX-X X.XX KV
>
40 mA