OPERATION
66
what conditions activate each pin. When a terminal becomes active the relay closes
thereby allowing the external voltage to operate an external device.
Pins 1 and 2 provide the PASS signal.
Pins 3 and 4 provide the FAIL signal.
Pins 5 and 6 provide the PROCESSING signal.
Pins 7, 8, and 9 are blank.
The following describes how the relays operate for each test condition.
PROCESSING - The relay contact closes the connection between pin (5) and pin (6)
while the instrument is performing a test. The connection is opened at the end of the
test.
PASS - The relay contact closes the connection between pin (1) and pin (2) after
detecting that the item under test passed all tests. The connection is opened when
the next test is initiated or the reset function is activated.
FAIL - The relay contact closes the connection between pin (3) and pin (4) after
detecting that the item under test failed any test. The connection is opened when the
next test is initiated or the reset function activated.
6.2 Signal Inputs of Remote I/O and Memory Programs
The HypotULTRA II or QUADCHEK II remote connector enables remote
operation of the TEST and RESET functions or allows the operator to select one of
three pre-programmed tests. When the remote function is (ON) the TEST switch on
the front panel will be disabled to prevent a test from being activated through the
test switch. A normally open momentary switch can then be wired across pins 3 and
5 to allow remote operation of the TEST function. A normally open momentary
switch can also be wired across pins 2 and 5 which allows remote operation of the
RESET function. For safety the front panel RESET switch remains active even
when a remote RESET switch is connected so that high voltage can be shut down
from either location.
The HypotULTRA II or QUADCHEK II also allow access to three MEMORY
PROGRAMS through the remote control connector. This gives the user the
capability to quickly change parameters and initiate a test remotely. The
HypotULTRA II or QUADCHEK II basically operates in a PLC mode by
responding to simple switch or relay contact closures. The built in memory
programs of the instrument are used to accomplish this. Three internal memory
programs can be accessed, by connecting terminals 7, 8 and 9 in different
combinations.
ACTIVATING MEMORY PROGRAM FUNCTIONS THROUGH THE
REMOTE CONNECTOR, SELECTS THE PROGRAM AND STARTS
THE TEST WHICH IS PREPROGRAMMED INTO THAT MEMORY
WARNING