OPERATION
41
2.2.4 Ramp Time setting
Advance the menu to the Ramp Time parameter. The display will show:
Ramp Time = XXX.X s
Range : 0.4 - 999.9
Use the Numeric keys to enter the Ramp Time setting, then press the ENTER key .
The program will store the Ramp Time setting and advance to the Dwell Time
parameter automatically. The unit is second and 0.1 second per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
2.2.5 Dwell Time setting
Advance the menu to the Dwell Time parameter. The display will show:
Dwell Time = XXX.X s
Range : 0.5 - 999.9 0 = Constant
Use the Numeric keys to enter the Dwell Time setting, then press the ENTER key.
The program will store the Dwell Time setting and advance to the Charge-LO
parameter automatically. The unit is second and 0.1 second per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
If the Dwell Time is set to
, the timer will continue to count to the maximum test
time then reset to 0 and start over automatically. The test will continue until a reset
is executed or a failure occurs.
2.2.6 Charge-LO
Advance the menu to the Charge-LO parameter. The display will show:
Charge-LO = XXX.X
m
A
<
TEST
>
to Auto Set
The Charge-LO function is used to check if the test cables are connected properly at
the beginning of a test. A capacitive DUT will draw charging current on the DC
Withstand Voltage test when the Output is activated. If the charging current was
lower then the setting, the test cables may not be connected properly.
This instrument can set the Charge-LO value manually or automatically. To set the
Charge-LO value manually, use the numeric keys to enter the Charge-LO current
setting and then press the ENTER key . The program will store the Charge-LO
setting and advance to the Ramp-HI parameter. The setting range of Charge-LO is
from 0.0 to 350.0
m
A. The unit is
m
A and 0.1
m
A per step.