OPERATION
49
Use the Numeric keys to enter the HI-Limit setting, then press the ENTER key. The
program will store the HI-Limit setting and advance to the LO-Limit parameter
automatically. The unit is m
W
and 1 m
W
per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
2.4.4 Low Limit Ground Resistance setting
Advance the menu to the LO-Limit parameter. The display will show:
LO-Limit = XXX m
W
Range : 0 - 600
Use the Numeric keys to enter the LO-Limit setting, then press the ENTER key.
The program will store the LO-Limit setting and advance to the Dwell Time
parameter automatically. The unit is m
W
and 1 m
W
per step.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
If the LO-Limit is set to
, the LO-Limit is disabled.
2.4.5 Dwell Time setting
Advance the menu to the Dwell Time parameter. The display will show:
Dwell Time = XXX.X s
Range : 0.5 - 999.9 0 = Constant
Use the Numeric keys to enter the Dwell Time setting, then press the ENTER key.
The program will store the Dwell Time setting and advance to the Frequency
selection parameter automatically. The unit is second and 0.1 second per step. The
display will show the elapsed time during the testing.
Press the EXIT key to exit from the setting mode to the operation mode if all
parameters have been set.
If the Dwell Time is set to
, the timer will continue to count to the maximum test
time then reset to 0 and start over automatically. The test will continue until a reset
is executed or a failure occurs.
2.4.6 Frequency Selection
Advance the menu to the Frequency parameter. The display will show:
Frequency = 60 Hz
<
ENTER
>
to Select
or
Frequency =
50 Hz
<
ENTER
>
to Select