QUICKSTART
31
QUICK START
This quick start guide assumes the operator has some familiarity with automated testing
and desires to use the
"default"
settings on the instrument. The default settings shown
will remain in memory unless you choose to override them with your own test program.
The instrument default settings that appear in memories 1-40 are as follows:
DEFAULTS
Function Parameter
Value
ACW Voltage
1240VAC
HI-Limit
10.00mA
LO-Limit
0.000mA
Ramp Time
1.0s
Dwell Time
1.0s
Frequency
60Hz
Arc Sense
5
Arc Fail
OFF
Scanner
OOOOOOOOOOOOOOOO
Connect
OFF
Continuity
(7500DT
only)
OFF
DCW Voltage
1500VDC
HI-Limit
3500
m
A
LO-Limit
0.0
m
A
Ramp Time
1.0s
Dwell Time
1.0s
Ramp-HI
OFF
Charge-LO
0.0
m
A
Arc Sense
5
Arc Fail
OFF
Scanner
OOOOOOOOOOOOOOOO
Connect
OFF
Continuity
(7500DT
only)
OFF
IR
Voltage
1000VDC
Charge-LO
0.000
m
A
HI-Limit
0M
W
LO-Limit
10M
W
Delay
1.0s
Scanner
OOOOOOOOOOOOOOOO
Connect
OFF
Содержание HypotULTRA II 7500DT