OPERATION
60
If the DUT current is well beyond the metering range of AC Withstand Voltage test
the instrument assumes that the failure is due to a short circuit, the display will show:
ACW
Short XXX.X s
MXX-X - . - - KV
>
40 mA
If the DUT current is well beyond the metering range of AC Withstand Voltage test
and an Arcing condition beyond the Arc Sense limit is indicated, the display will
show:
ACW Breakdown XXX.X s
MXX-X X.XX KV
>
40 mA
If the DUT current falls below the LO-Limit of AC Withstand Voltage test the
display will show:
ACW
LO-limit XXX.X s
MXX-X X.XX KV XX.XX mA
If the DUT current is within the metering range of the AC Withstand Voltage test
and an Arcing current exceeds the Arc-Sense limit and the Arc function is set to ON,
then an Arc failure has occurred and the display will show:
ACW
Arc-Fail
XXX.X s
MXX-X X.XX KV XX.XX mA
If the GFI threshold is exceeded during of the AC Withstand Voltage test, the
display will show:
ACW GND-Fault XXX.X s
MXX-X X.XX KV XX.XX mA
When the DUT passed the AC Withstand Voltage test, when the test process is
complete the display will show:
ACW
Pass
XXX.X
s
MXX-X X.XX KV XX.XX mA
5.3 The followings are the displayed messages for the DC Withstand Voltage test:
If the test in process is Aborted with the RESET button or remote control, the
display will show:
DCW
Abort XXX.X s
MXX-X X.XX KV XXXX
m
A