C
SP-4430 ● 5-05
4-1
お手入れの概要
Chapter 5 TROUBLESHOOTING
4-1-1
お手入れの目安
( 7 〜 9 ) : Single type
( M ) : Multi type
Channel with expired Reagent Strips in
either case.
The clock is not correctly adjusted.
The power was not connected for a long
time (battery is dead).
●
●
Readjust the date and time.
(see “3-6 Built-in Clock Adjustment”)
●
E31
E31 Wrong date/time
■■■■■■■■■■■■
OK(ENTER)
●
●
●
●
Use a new Reagent Strip.
Adjust the clock.
Insert a new magnetic card.
●
●
●
E32
Description and error message
Conditions and causes
Remedy
E32 Invalid strips
(
■■
7
■■■■■
M)
■
OK(ENTER)
External light entered because the table
cover or maintenance cover was slightly
opened.
●
Close the table cover or the
maintenance cover firmly.
Change the direction of the
analyzer.
●
●
E33
E33 Stray light
■■■■■■■■■■■■
OK(ENTER)
( 7 〜 9 ) : Single type
Channel which was misread.
The table is dirty.
The bar code was not correctly read
due to displacement or bending of the
Reagent Strip.
Foreign matter is attached to the Single
Reagent Strip.
●
●
●
Remove dust from the table.
Reset the Single Reagent Strip
correctly.
Confirm that foreign matter is not
attached to the Single Reagent
Strip.
●
●
●
E34
E34 BAR misread
(987
■■■■■■
)
■
OK(ENTER)
The setting (such as baud rate) of the
hand-held bar code reader is incorrect.
The hand-held bar code reader is not
correctly connected.
●
●
Confirm the setting of the hand-
held bar code reader (such as baud
rate).
Connect the hand-held bar code
reader correctly.
●
●
E35
E35 BCR error
■■■■■■■■■■■■
OK(ENTER)
Battery is exhausted.
●
Charge the battery.
●
E36
E36 Battery error
■■■■■■■■■■■■
OK(ENTER)
Wear protective gloves in order to avoid exposure
to pathogenic microbes.
WARNING
The Reagent Strip has expired.
The clock is not correctly adjusted.
The clock is not correct because the
power was not connected for a long
time.
When using a Reagent Strips from new
lot, the corresponding magnetic card to
the lot of the was not inserted.
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...