Chapter 2 MEASUREMENT
2-3-4
検体の準備
Set the samples.
7.
Open the protective cover and place
the uncapped centrifuge tube in
position.
Close the cover.
●
●
■
Measurement of uncentrifuged whole blood.
NOTE
Make sure the sample tubes are
uncapped, otherwise it may
cause breakage of centrifuge
tube and the Nozzle.
Set a sample after removing the
cap.
Uncap sample tubes and place them
to the port of the multi-rack.
●
■
Measurement of serum or plasma, or centrifuged samples.
NOTE
Make sure the sample tubes are
uncapped, otherwise it may
cause breakage of the Nozzle.
NOTE
Make sure the protective cover is
closed properly.
If the protective cover remains
open, it may become troubled
and correct measurement results
may not be obtained.
NOTE
Make sure the protective cover is
closed properly.
If the protective cover remains
open, it may cause a trouble.