C
5-02 ● SP-4430
5-1
Error Messages
Chapter 5 TROUBLESHOOTING
4-1-1
お手入れの目安
When an error occurs, an alarm sounds and the error message appears
on the screen.
The error message is automatically printed in order to preserve the
error record (The following errors are only displayed, and not printed).
( T ) : Trouble with the table cover.
( M ) : Trouble with the maintenance cover.
●
The following errors are not
printed when the magnetic
card is inserted.
E15
No MEAS data
E21
Card misread
E24
Wrong card
E25
Wrong stripe
The table cover was opened during
measurement or warm-up.
The maintenance cover was opened
during measurement or warm-up.
The table cover is not properly closed
(Foreign matter is jammed).
The measurement was started without
closing the maintenance cover.
●
●
●
●
Close the table cover correctly
(Remove the foreign matter).
Set the maintenance cover in the
correct position.
●
●
E02
Description and error message
Conditions and causes
Remedy
E02 Cover open
■■■■■■■■■■■■
OK(ENTER)
Power failure has occurred during
measurement.
Power was turned off during
measurement.
The power cable was unplugged during
measurement.
●
●
●
The last measurement was invalid.
Restart measurement.
●
E03
E03 Power down
■■■■■■■■■■■■
OK(ENTER)
A light source is deteriorated.
●
If the same trouble occurs, contact
your distributor.
●
E04
E04 LED unstable
■■■■■■■■■■■■
OK(ENTER)
The white and black plate or window
plate is dirty.
●
Clean the white and black plate or
window plate ( see “4-2-1 Cleaning
the Reagent Table” and “4-3-2
Cleaning the Optical Window”).
●
E05
E05 Optical error
(9
■■
6
■■■
2
■
)
■
OK(ENTER)
(1 〜 9) : Channels with abnormalities
Reagent Strip is not set.
The bar code was not correctly read
due to displacement or bending of the
Reagent Strip.
The Reagent Strip is not set in the
channel necessary for calibration.
●
●
●
Set Reagent Stripes correctly.
When calibrating, set the necessary
numbers.
●
●
E11
E11 No strips
(
■■■■■■■■
M)
■
OK(ENTER)
( 7 〜 9 ) : Single type
( M ) : Multi type
Channel with no Reagent Strip in either case.
Display is made only in calibration. Usually
nothing is displayed during measurement.
Wear protective gloves in order to avoid exposure
to pathogenic microbes.
WARNING
Wear protective gloves in order to avoid exposure
to pathogenic microbes.
WARNING
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...