C
SP-4430 ● 4-11
4-1
お手入れの概要
Chapter 4 MAINTENANCE
4-1-1
お手入れの目安
Clean the Optical Window.
3.
Remove the rubber cap on the
bottom of the analyzer.
●
Turn over the analyzer carefully so
that the right side (to which small
legs are attached) faces down.
Do NOT hold the Table Cover while
turning the analyzer.
●
Wipe out stains and dust adhered to
the Optical Window (transparent
glass plate) with a cotton swab
moistened with distilled water.
Wipe out remained cotton swab
fibers and moisture of the Optical
Window with a new dry cotton
swab. Check it the Optical Window
is clean and dust-free. Use a
flashlight to make the checking
easier.
●
●
NOTE
Do NOT damage the connecting
part of the Table Cover.
Be careful not to tip over the
analyzer to avoid damage.
Содержание SPOTCHEM EZ SP-4430