Chapter 2 MEASUREMENT
Measuring the Hb
When measuring Hb, dilute whole blood
samples 6 times with distilled water.
Prepare a Whole blood Sample Tube (orange
cap).
Uncap the tube and pipette the whole blood
sample into the tube.
Cap the tube tight.
Invert the sample several times ( see “2-14
Inverting a sample”).
Prepare a Serum Sample Tube(blue cap).
Pipette 200
μL
of distilled water.
Pipette 40
μL
of the prepared whole blood
sample into the above Serum Sample Tube.
Cap the tube tight.
Invert the sample several times again.
●
●
●
●
●
●
●
●
Whole blood sample tube
Orange cap
distilled water
(
200
)
Blue cap
Serum sample tube
distilled water
(
200
)
Remove any air bubbles or skin on the surface of the
sample.
They may cause incorrect measurement results.
IMPORTANT