C
SP-4430 ● 4-13
4-1
お手入れの概要
Chapter 4 MAINTENANCE
4-3-3
Cleaning the Nozzle
With numerous measurements, sample blood residue may stay inside the
Nozzle and cause clogging.
Clean the Nozzle once every 10,000 measurements or once a year.
Remove the cover.
2.
Remove the top cover by sliding it to
the left.
●
Move the Nozzle to stand-by position.
1.
Press [2] key on the MAIN MENU.
SUB MENU 1/2 is displayed.
Press [3] key .
The Table Cover opens, and the
Reagent Table and Centrifuge-
equipped Multi Rack slide forward.
The Nozzle Driving Unit moves to
the left end in the analyzer.
Turn OFF the power.
●
●
●
Requirements
Nozzle cleaning wire, Tweezers,
Tissue paper and Protective gloves
Wear protective gloves to prevent exposure to pathogenic microbes.
WARNING
Discard used samples, tips and protective gloves separately from general
waste according to local regulations on biohazardous waste.
CAUTION
1 . R e s u l t s 2 . P A R AM
3 . M a i n t e n a n c e
( 1/2 )
1 . M e a s u r e 2 . S u b m e n u
3 . C a l i b r a t e
( 1/1 )
R e a d y .
P l e a s e
t u r n o f f .
Make sure the power is OFF before moving on to the next procedure.
Otherwise, it may cause injury by contacting with the driving unit of the
analyzer.
CAUTION
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...