SP-4430 ● 2-35
2-5
校正
Chapter 2 MEASUREMENT
2-5-1
校正の概要
Insert the Reagent Strips with the
same items, starting from the S1
channel on the Reagent Table.
If the number of calibration is set as
more than 3 for 2-point calibration,
calibration should be sperately
performed twice. In this case , set 3
Reagent Strips for the first
calibration.
(ex.: When the number of calibration
is 4, set 3 Reagent Strips for the first
calibration and 1 for the second .)
●
With Multi-type Reagent Strips,
1-point calibration cannot be
performed.
▲
●
Calibration with Multi-type Reagent Strips
Prepare Multi-type Reagent Strips.
Prepare twice as many Multi-type
Reagent Strips as the number of the
times calibration that will be
performed.
Set the Strip on the Reagent Table
from the right hand side.
●
●