C
5-04 ● SP-4430
4-2
毎日のお手入れ
Chapter 5 TROUBLESHOOTING
4-2-1
試験片テーブルのお手入れ
PR : Prozone error
L 1 : Large difference between the measured
and the displayed values of calibrator L.
H 1 : Large difference between the measured
and the displayed values of calibrator H.
L 2 : Large variations between the measured
values of calibrator L.
H 2 : Large variations between the measured
values of calibrator H.
LH : No difference between the measured
value of calibrator L and that of calibrator
H.
( 1 〜 6 ) : Channel with an error.
The display is shown only in multi
calibration, and nothing is displayed in
single calibration.
The magnetic card was caught when
inserted.
The head of magnetic card reader is
dirty.
●
●
Insert the magnetic card again.
Clean the head of the magnetic
card reader.
●
●
E21
Description and error message
Conditions and causes
Remedy
E21 Card misread
■■■■■■■■■■■■
OK(ENTER)
Wrong magnetic card was used (a
calibration card was inserted when the
display showed Reagent Card reading
or vice-versa).
●
Insert a correct magnetic card.
●
E24
E24 Wrong card
■■■■■■■■■■■■
OK(ENTER)
The same stripe was inserted.
●
Insert a different stripe.
●
E25
E25 Wrong stripe
■■■■■■■■■■■■
OK(ENTER)
During calibration, the difference of the
measured value and the display of
calibrator is extremely large.
The variations of the measured values
are extremely large in calibration.
Calibrator L and H might be set
oppositely in calibration.
Incorrect adjustment of calibrator.
●
●
●
●
Perform measurement again.
●
E30
E30 Abnormal data
(L1
■■■■■■
3)
■
OK(ENTER)
Cuvette: when a curette is used for liquid
level detection.
CNTRFG: When a centrifuge vessel is used.
A sample vessel is misplaced.
There is no sample or an insufficient
amount of sample.
A centrifuge tube was used in
calibration by calibrator kit.
●
●
●
Set the sample correctly.
Confirm the amount of the sample.
Be sure to use cuvettes in
calibration by calibrator kit.
●
●
●
E17
E17 No sample
(CNTRFG
■■■
)
■
OK(ENTER)
Wear protective gloves in order to avoid exposure
to pathogenic microbes.
WARNING
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...