2-3
測定結果
Chapter 2 MEASUREMENT
2-4-1
通常測定
Press [0] key on the Standby screen.
The Centrifuge selecting screen is
displayed.
[On] is always displayed at the
beginning.
To switch [On] and [Off] of
centrifuge, press [0] key on the
Centrifuge selecting screen.
●
●
■
Remeasurement (or additional measurement) with a centrifuge.
When measuring the once-measured sample with a centrifuge, such measurement can
be done at a reduced centrifugal time.
When performing remeasurement or additional measurement , sample
quantity may be short. Shortage of sample amount may lead to incorrect
results due to sucking of blood cells or trouble because of abnormal rotation
of centrifuge. Make sure that the remaining sample amount is enough, and
perform remeasurement or additional measurement.
IMPORTANT
To perform the measurement on the screen of [No centrifugation], be sure to
use the centrifuged sample. If uncentrifuged sample is used, correct
measurement result may not be obtained.
IMPORTANT
Every time when [0] key is
pressed on the Centrifugation
selecting screen, [On], and [Off]
are displayed alternately.
▲
●
Set the Multi-type Reagent Strips.
Set the Reagent Strips.
8.
Multi-type Reagent Strip
Open the aluminum foil package
from the place written “
▲
Open here
▲
” until two-thirds of the Reagent
Strip is exposed.
●
In Normal Measurement, the following three settings are available.
●
Multi-type Reagent Strips and Single type Reagent Strip
●
Multi-type Reagent Strips only.
●
Single-type Reagent Strips only.
Choose one depending on the item to measure.
S t a n d b y
2000ー01ー19
I D( 1 ) I N F O( 2 )
S t a n d b y
2000ー01ー19
CNTRFG=[ ON ]
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...