Chapter 1 INTRODUCTION
Before setting up the
analyzer, read the
following notes and
always take proper
safety precautions.
CAUTION
Do NOT set up the analyzer where
chemicals are stored nearby, or where
corrosive gases or electrical noise are
generated. They may damage the analyzer,
and malfunctions and/or injury may occur.
Correct measurement results may not be
obtained.
●
Avoid exposing the analyzer to direct
sunlight, condensation and wind.
Otherwise, correct measurement results
may not be obtained, and may cause
deformation or malfunction of the
analyzer.
●
To connect the analyzer to external
devices, be sure to use proper cables to
avoid electric shock and/or fire. For
details, contact your distributor.
●
temperature
10
〜
30
℃
humidity
20
〜
80
%
Place the analyzer in a room at
temperatures between 10
℃
and 30
℃
with humidity between 20% and 80%.
Otherwise, correct measurement results
may not be obtained.
●
Make sure that the room is well ventilated when a large
amount of carbon dioxide is generated (when gas
stoves, oil heaters and instantaneous water heaters are
used in the room). This is because the pH of the
Reagent Strips, which use an alkaline reaction reagent,
decreases under the influence of carbon dioxide and
correct measurement data may not be obtained.
●
Be careful not to be put hands under the
analyzer.
●
Содержание SPOTCHEM EZ SP-4430
Страница 1: ...b SPOTCHEM EZ SP 4430 Operating Manual ...
Страница 8: ...C 07 SP 4430 ...
Страница 9: ...b ...
Страница 25: ...C 1 16 SP 4430 ...
Страница 26: ...b ...
Страница 31: ...e SP 4430 2 05 1 4 装置の設置 Chapter 2 MEASUREMENT 2 1 4 校正について The flowchart of calibration ...
Страница 70: ...C 2 44 SP 4430 ...
Страница 71: ...b ...
Страница 74: ...C SP 4430 3 03 1 4 装置の設置 Chapter 3 SUB MENU 1 4 3 設設置後の始動と終了 ...
Страница 97: ...C 3 26 SP 4430 ...
Страница 98: ...b ...
Страница 117: ...b ...
Страница 129: ...b 5 12 SP 4430 Chapter 5 TROUBLESHOOTING ...
Страница 130: ...b ...
Страница 139: ...b 83 00139F ...