Page
88
Command Protocol
WCOX
Read code
226
characters.
This command outputs all data of a selected code.
Parameters are 7 figures with "-" sign at MSB when negative
value and without decimal point. If code number is replaced
with space codes (20h), the current code is output.
Code
number
Output item order
Length
Code number
RCOD + 4
Code name
12 + 3 spaces
Target 7
Hi 7
Lo 7
HiHi 7
LoLo 7
Zero band
7
Full 7
Preset tare
7
Target count
7
Total count
7
OK count
7
NG count
7
Hi count
7
Lo count
7
HiHi count
7
LoLo count
7
Foreign matter count
7
Duplication count
7
Crush count
7
Maximum 7
Minimum 7
Average 7
Standard deviation
σ
n-1
7
Population standard deviation
σ
n
7
Total weight
9
RCOD0034
C
R
L
F
RCOD0034,12345678...,123456789
Computer
AD-4404
C
R
L
F
Содержание Check Weighing Indicator AD-4404