AD-4404 Page
139
Category address: [Function] - [Set Function] - [Sequence] - [Zero track]
Parameters for Dynamic Zero Tracking Function
Category
address
symbol
Name Descriptions
Range
and
choices
Default
5q f-71
Zero tracking
back timer
The time to stop zero tacking before
weighing the article.
0.0 to
5.0
s
1.0
s
5q f-72
Tracking
average timer
Average time.
0.0 to
5.0
s
1.0
s
5q f-73
Tracking MIN
interval
Minimum interval between weighings.
If next weighing performs before the
timer is up, zero tracking is not
performed.
0.0 to
99.9
s
10.0
s
5q f-74
Tracking
amplitude
Tolerance value to perform zero tacking.
If absolute average value of zero
tracking is above this parameter, zero
tracking is not performed.
0.0 to
99.9
d
0.0
d
Refer to "8.1.2. Dynamic Zero Tracking Function".
s
: second,
d
: digit
Position sensor
Eval delay timer
5q f-42
Judgement
Chattering timer
Gross
Detection
5q f-72
, Zero tracking average timer
It is time to perform zero tacking.
5q f-71
, Zero tacking back timer
It is time to stop zero tacking before
weighing the article.
Judgement
Zero tracking average timer
Zero tracking MIN interval
5q f-73
, Zero tracking MIN interval
If the article is detects within time of
5q f-73
, zero tracking is not
attained.
Detection
Zero tracking is calculated at this time
0
5q f-72
,
Zero tracking average timer
5q f-74
,
Zero tracking amplitude
5q f-43
5q f-72
5q f-73
Содержание Check Weighing Indicator AD-4404
Страница 1: ...Check Weighing Indicator WM PD4000489...